One of our oldest gas mixers is currently back at our facility for service, maintenance and calibration.
Even though the system has already been in operation for several years, it is still one of the most comprehensive gas mixing systems we have built.
The unit was designed as a mobile system and is used by different customers for a wide range of applications. Its configuration is therefore equally flexible – from gas mixtures in the ppm range to conventional process gas.
Systems like this clearly demonstrate the importance of regular maintenance and calibration. At the time this generation was developed, the automatic calibration function that is standard in our systems today had not yet been integrated. The sensors of the gas analysis system therefore need to be checked and recalibrated during servicing to ensure that the measured gas composition can continue to be assessed reliably.
With our current systems, this process is much more automated. The sensors of the integrated gas analysis can be automatically checked and calibrated with zero and calibration gas at freely configurable intervals.
This provides benefits beyond the analysis itself. Changes in the behaviour of the mass flow controllers over their operating lifetime also become visible through the measured gas composition. If the actual composition deviates from the target value, the mixing recipe can be adjusted accordingly.
This makes gas analysis an important tool for the long-term quality assurance of the entire gas mixing system.
For us, servicing systems like this is always interesting for another reason as well: it not only demonstrates the longevity of the equipment, but also shows very clearly how our technology has evolved over the years.